Jjg443-2006 verification regulation of fuel dispenser stipulates the error of fuel dispenser as follows:
1 measurement performance requirements
1.1 maximum allowable error of fuel dispenser
1.1.1 the maximum allowable error of fuel dispenser is ± 0.3%, and its repeatability shall not exceed 0.15%.
1.1.2 the minimum measured maximum allowable error of the fuel dispenser shall not exceed ± 0.5%, and its measurement repeatability shall not exceed 0.21%.
1.2 maximum allowable error of flow measurement converter
The maximum allowable error of flow measurement converter should not exceed ± 0.2%, and its repeatability should not exceed 0.1%.
1.3 flow ratio of fuel dispenser
The ratio of the maximum flow rate Qmax and the minimum flow rate qmin of the newly produced fuel dispenser shall not be less than 10:1.
1.4 flow ratio of flow measurement converter
The ratio of the maximum flow Qmax and the minimum flow qmin of the flow measurement converter shall not be less than 10:1.
1.1 minimum measured value of fuel dispenser
If the maximum flow rate is not more than 60L / min, the minimum measured flow rate shall not exceed 5L.
1.6 minimum volume variable of fuel dispenser
The minimum volume variable of fuel dispenser shall not be greater than 0.02l